GW Instek GPT-9603 100VA AC/DC/IR Withstanding Voltage Tester
|
Item Number: GPT-9603
Manufacturer: GW Instek
Part Number: GPT-9603
* Whole number only
|
|
GW Instek GPT-9603 100VA AC/DC/IR Withstanding Voltage Tester
The GPT-9603 safety tester is a 3-in-1 instrument for voltage safety withstand testing (hipot) and insulation resistance testing. Perform AC hipot tests to 5000 volts AC or 6000 VDC. The GPT-9603 also performs insulation resistance testing to 2000 M-Ohms. The bright blue Dot Matrix LCD gives the test engineer all they need at their finger tips to meet IEC, EN, UL, ETL, CSA, GB, JIS and other safety standards. Custom calibration certificates are available to help the end user prove traceability and accuracy to their standards inspection body or organization.
Features
- 100VA AC Test Capacity
- 240x48 Ice Blue Dot matrix LCD
- Auto W-I or I-W Mode
- ARC Detection
- True RMS Current Measurement
- Zero Crossing Turn-on Operation
- Safety INTERLOCK function
- PWM amplifier to enhance the power efficiency and reliable testing
- Automatically Switching Input Source for Universal Input Voltage
- Interface : Remote terminal and Signal I/O
Technical Specifications
AC WITHSTANDING (All Models)
|
|
Output-Voltage Range
|
0.10kV~ 5.00kV ac
|
|
Output-Voltage Resolution
|
10V
|
|
Output-Voltage Accuracy
|
± (1.5% of setting + 2 counts) with no load
|
|
Maximum Rated Load
|
100 VA (5kV/20mA)
|
|
Maximum Rated Current
|
20mA (0.5kV < V ≦ 5kV)
5mA (0.1kV ≦ V≦ 0.5kV)
|
|
Output-Voltage Waveform
|
Sine wave
|
|
Output-Voltage Frequency
|
50 Hz / 60 Hz selectable
|
|
Voltage regulation
|
± (1.5% + 2 counts) [full load → no load]
|
|
Voltmeter Accuracy
|
± (1.5% of reading + 2 counts)
|
|
Current Measurement Range
|
0.01mA~20.0mA
|
|
Current Best Resolution
|
0.01mA / 0.1mA
|
|
Current Measurement Accuracy
|
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
|
|
Current Judgment Accuracy
|
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
|
|
Window Comparator Method
|
Yes
|
|
ARC Detect
|
Yes
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
ON
|
DC WITHSTANDING (GPT-9602 & GPT-9603)
|
|
Output-Voltage Range
|
0.10kV~ 6.00kV dc
|
|
Output-Voltage Resolution
|
10V
|
|
Output-Voltage Accuracy
|
± (1.5% of setting + 2 counts) with no load
|
|
Maximum Rated Load
|
25W (5kV/5mA)
|
|
Maximum Rated Current
|
6mA (0.5kV < V ≦ 6kV)
2mA (0.1kV ≦ V≦ 0.5kV)
|
|
Voltage regulation
|
± (1.5% + 2 counts) [full load → no load]
|
|
Voltmeter Accuracy
|
± (1.5% of reading + 2 counts)
|
|
Current Measurement Range
|
0.01mA~6.00mA
|
|
Current Best Resolution
|
0.01mA
|
|
Current Measurement Accuracy
|
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
|
|
Current Judgment Accuracy
|
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
|
|
Window Comparator Method
|
Yes
|
|
ARC Detect
|
Yes
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
ON
|
INSULATION RESISITANCE (GPT-9603 & GPT-9612)
|
|
Output-Voltage
|
50V, 100V, 250V, 500V, 1000V dc
|
|
Output-Voltage Accuracy
|
± (3.0% of setting + 1 count) [no load]
|
|
Resistance Measurement Range
|
1MOhm~ 2000M-Ohm
|
|
|
Test voltage
|
Measurable Range
|
Accuracy
|
|
|
50V/100V/250V
|
1~ 50M-Ohm
51~ 2000M-Ohm
|
±(5% of reading + 2M-Ohm)
±(10% of reading + 2M-Ohm)
|
|
|
500V/1000V
|
1~ 500M-Ohm
501~ 2000M-Ohm
|
±(5% of reading + 2M-Ohm)
±(10% of reading + 2M-Ohm)
|
|
Window Comparator Method
|
Yes
|
|
Output Impedance
|
600k-Ohm
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
OFF (fixed)
|
TEST MODE*
|
|
SINGLE
|
ACW, DCW, IR
|
|
AUTO
|
AC-IR, IR-AC, DC-IR, IR-DC
|
INTERFACE
|
|
Remote Terminal (Front)
|
Standard
|
|
SIGNAL IO
|
Standard
|
General
|
|
DISPLY
|
240 x 48 Ice Blue dot matrix LCD
|
|
POWER SOURCE
|
AC 100V-120V/220V-240V ±10%, 50/60Hz
|
|
POWER COMSUMPTION
|
400VA max.
|
|
DIMENSIONS & WEIGHT
|
330(W) x 148(H) x 385(D) mm
approx.9kg(Max)
|
* The available “Test Mode” depends on selected model
IEC, EN, UL, ETL, CSA, GB, JIS and other
|
The GW Instek GPT-9601 provides voltage breakdown testing to 5 KV AC. The GPT-9601 is the ideal hipot for quick AC only electrical voltage withstand safety testing.
|
The GW Instek GPT-9612 provides AC hipot output up to 5000 Volts AC and insulation resistance (IR) tests to 2000 Meg Ohms.
|
The GW Instek GPT-9602 provides a AC and DC voltage breakdown testing up to 5000 VAC and 6000 VDC.
| |
Product Reviews
|
|