GW Instek GPT-9601 5KV AC Hipot Voltage Breakdown Tester
|
Item Number: GPT-9601
Manufacturer: GW Instek
Part Number: GPT-9601
* Whole number only
|
|
GW Instek GPT-9601 5KV AC Withstanding Voltage Tester
The GPT-9601 safety tester is capable of performing AC withstanding test up to 5000 Volts AC. The GPT-9600 series is equipped with the high-efficiency PWM amplifier. The bright blue Dot Matrix LCD gives the test engineer all they need at their finger tips to meet IEC, EN, UL, ETL, CSA, GB, JIS and other safety standards. Custom calibration certificates are available to help the end user prove traceability and accuracy to their standards inspection body or organization.
Features
- 100VA AC Test Capacity
- 240x48 Ice Blue Dot matrix LCD
- Auto W-I or I-W Mode (Modesl with IR feature only)
- ARC Detection
- True RMS Current Measurement
- Zero Crossing Turn-on Operation
- Safety INTERLOCK function keeps the test area safe
- PWM amplifier to enhance the power efficiency and reliable testing
- Automatically Switching Input Source for Universal Input Voltage
- Interface : Remote terminal and Signal I/O
Technical Specifications
AC WITHSTANDING (All Models)
|
|
Output-Voltage Range
|
0.10kV~ 5.00kV ac
|
|
Output-Voltage Resolution
|
10V
|
|
Output-Voltage Accuracy
|
± (1.5% of setting + 2 counts) with no load
|
|
Maximum Rated Load
|
100 VA (5kV/20mA)
|
|
Maximum Rated Current
|
20mA (0.5kV < V ≦ 5kV)
5mA (0.1kV ≦ V≦ 0.5kV)
|
|
Output-Voltage Waveform
|
Sine wave
|
|
Output-Voltage Frequency
|
50 Hz / 60 Hz selectable
|
|
Voltage regulation
|
± (1.5% + 2 counts) [full load → no load]
|
|
Voltmeter Accuracy
|
± (1.5% of reading + 2 counts)
|
|
Current Measurement Range
|
0.01mA~20.0mA
|
|
Current Best Resolution
|
0.01mA / 0.1mA
|
|
Current Measurement Accuracy
|
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
|
|
Current Judgment Accuracy
|
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
|
|
Window Comparator Method
|
Yes
|
|
ARC Detect
|
Yes
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
ON
|
DC WITHSTANDING (GPT-9602 & GPT-9603)
|
|
Output-Voltage Range
|
0.10kV~ 6.00kV dc
|
|
Output-Voltage Resolution
|
10V
|
|
Output-Voltage Accuracy
|
± (1.5% of setting + 2 counts) with no load
|
|
Maximum Rated Load
|
25W (5kV/5mA)
|
|
Maximum Rated Current
|
6mA (0.5kV < V ≦ 6kV)
2mA (0.1kV ≦ V≦ 0.5kV)
|
|
Voltage regulation
|
± (1.5% + 2 counts) [full load → no load]
|
|
Voltmeter Accuracy
|
± (1.5% of reading + 2 counts)
|
|
Current Measurement Range
|
0.01mA~6.00mA
|
|
Current Best Resolution
|
0.01mA
|
|
Current Measurement Accuracy
|
±(2.0% of reading + 5 counts) when HISET < 1.00mA
±(2.0% of reading + 3 counts) when HISET ≧ 1.00mA
|
|
Current Judgment Accuracy
|
±(3.0% of setting + 5 counts) when HISET < 1.00mA
±(3.0% of setting + 3 counts) when HISET ≧ 1.00mA
|
|
Window Comparator Method
|
Yes
|
|
ARC Detect
|
Yes
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
ON
|
INSULATION RESISITANCE (GPT-9603 & GPT-9612)
|
|
Output-Voltage
|
50V, 100V, 250V, 500V, 1000V dc
|
|
Output-Voltage Accuracy
|
± (3.0% of setting + 1 count) [no load]
|
|
Resistance Measurement Range
|
1MOhm~ 2000M-Ohm
|
|
|
Test voltage
|
Measurable Range
|
Accuracy
|
|
|
50V/100V/250V
|
1~ 50M-Ohm
51~ 2000M-Ohm
|
±(5% of reading + 2M-Ohm)
±(10% of reading + 2M-Ohm)
|
|
|
500V/1000V
|
1~ 500M-Ohm
501~ 2000M-Ohm
|
±(5% of reading + 2M-Ohm)
±(10% of reading + 2M-Ohm)
|
|
Window Comparator Method
|
Yes
|
|
Output Impedance
|
600k-Ohm
|
|
RAMP (Ramp-Up Time)
|
0.1s fixed
|
|
TIMER (Test Time)
|
OFF, 1s~180s
|
|
GND
|
OFF (fixed)
|
TEST MODE*
|
|
SINGLE
|
ACW, DCW, IR
|
|
AUTO
|
AC-IR, IR-AC, DC-IR, IR-DC
|
INTERFACE
|
|
Remote Terminal (Front)
|
Standard
|
|
SIGNAL IO
|
Standard
|
General
|
|
DISPLY
|
240 x 48 Ice Blue dot matrix LCD
|
|
POWER SOURCE
|
AC 100V-120V/220V-240V ±10%, 50/60Hz
|
|
POWER COMSUMPTION
|
400VA max.
|
|
DIMENSIONS & WEIGHT
|
330(W) x 148(H) x 385(D) mm
approx.9kg(Max)
|
* The available “Test Mode” depends on selected model
|
The GW Instek GPT-9612 provides AC hipot output up to 5000 Volts AC and insulation resistance (IR) tests to 2000 Meg Ohms.
|
The GW Instek GPT-9602 provides a AC and DC voltage breakdown testing up to 5000 VAC and 6000 VDC.
|
The GW Instek GPT-9603 provides AC (5000 Volt) and DC (6000 Volt) and insulation resistance (IR) tests to 2000 Meg Ohms.
| |
Product Reviews
|
|